digitalYEWFLO Vortex Flow Meter

The digitalYEWFLO vortex flow meter is accurate and stable, even in harsh process conditions, and has a highly reliable and robust design that delivers improvements in plant efficiency and reduced operating costs.

0
0
0

The digital YEWFLO vortex flow meter combines the field proven sensor and body assembly used in more than 400,000 units installed worldwide with a unique and powerful combination of digital technology that includes spectral signal processing (SSP), a Yokogawa innovation. The digitalYEWFLO vortex flow meter is accurate and stable, even in harsh process conditions, and has a highly reliable and robust design that delivers improvements in plant efficiency and reduced operating costs


Standard

Multi-Variable

Reduced Bore

Model & code

DY Vortex Flowmeter

Optional code:/MV

Option code:/R1,/R2

Fluid to be measured

Liquid, gas, steam (avoid multiphase flow and sticky fluids)

Nominal size

15mm  300mm
(400mm: special)

25mm  200mm

25mm  200mm

Accuracy

Liquid: ±1.0% of reading (20000≤Re≤1000)
Liquid: ±0.75% of reading (D×1000≤Re)

Liquid: ±1.0% of reading
(20000≤Re≤1000)

Gas, steam: ±1.0% of reading (flow velocity less than 35m/s)
Gas, steam: ±1.5% of reading (flow velocity 35m/s 
 80m/s)

Output signal

Simultaneous output (both analog and transistor contact output available)

Process
temperature
range

-29 deg C - 260 deg C
(standard)

-29 deg C - 260 deg C
(standard)

-29 deg C - 260 deg C
(standard)

-196 deg C100 deg C
(option: cryogenic version)

N/A

N/A

-29 deg C450 deg C
(option: high process
temperature version)

N/A

-29 deg C450 deg C
(option: high process
temperature version)

Process
pressure
limit

-0.1MPa (-1kg/cm2) to flange rating

Ambient temperature

-29 to 85 deg C (Integral type / No indicator)
-29 to 80 deg C (Integral type / with indicator)
-29 to 85 deg C (Remote type detector)
-40 to 85 deg C (Remote type converter / No indicator)
-30 to 80 deg C (Remote type converter / with indicator)

Ambient
humidity

5100%RH( Non-condensing

Mounting

Integral type and Remote type detector: Flange mounting or wafer mounting Remote type converter: 2 inch pipe mounting
JIS10/20/40, ANSI150/300/600/900, JPI150/300/600, DIN PN10/16/25/40

Electrical connection

JIS G1/2 female, ANSI 1/2NPT female, ISO M20×1.5 female

Electrical classification

JIS flameproof, FM explosion proof/intrinsically safe
ATEX (CENELEC flameproof/intrinsically safe/IEC type n)
CSA explosion proof/intrinsically safe: Dual Seal
SAA flame proof/intrinsically safe

Material

Body: SCS14A cast stainless steel (equivalent to SUS316, CF8M)
Shedder bar: duplex stainless steel (option:anti-corrosion version)
Gasket: SUS316 stainless steel with polytetrafluoroethylene (Teflon) coating
Converter housing and case cover: aluminum alloy or ASTM CF8M (equiv. JIS SCS14A) (Option: /E1)

Applicable Standard

*EMC Conformity Standards: 
  EN61326-1 Class A, Table 2 (For use in industrial locations), EN61326-2-3 
  EN55011 Class A Group 1 
*Pressure Equipment Directive: 
  Notified Body Identification Number 0038 Module H




Carton or wooden box