The Deltapilot FMB52 is a cable version including a fixed process connection. This device with the Contite measuring cell is typically used in the process and environmental industries. It is made for level measurement in liquid and paste-like media in open or closed containers and is unaffected by possible foam formation. For use in SIL2 safety systems.
Measuring Principle | Hydrostatic |
Characteristic | Pressure transmitter for pressure or hydrostatic level measurement with flush mounted metallic CONTITE measuring cell: |
- hermetically welded |
- absolute condensation tight |
- climate-resitant |
- long-term stable |
- Lowest temperature influences |
|
Cable version |
Supply voltage | 4...20 mA HART |
10,5...45 VDC (Non Ex): |
Ex ia: 10,5...30 VDC |
PROFIBUS PA: |
9...32 VDC (Non Ex) |
FOUNDATION Fieldbus: |
9...32 VDC (Non Ex) |
|
Reference Accuracy | Standard 0,2% |
Optional 0,1% |
Long term stability | 0,05% of URL/year |
Process temperature | PE cable: -10°C…70°C / 14°F…158°F |
FEP cable: -10°C…80°C / 14°F…176°F |
Ambient temperature | -40°C...85°C |
(-40°F...185°F) |
Measuring cell | 100 mbar...10 bar |
(1.5 psi...150 psi) |
Smallest calibratable span | 10 mbar (1.45 psi) |
Vacuum resistance | 0 mbar abs. |
Max. Turn down | 100:01:00 |
Max. overpressure limit | 40 bar (600 psi) |
Process connection | Thread: |
G1 1/2, MNPT1 1/2 |
Flange: |
DN40...DN100, |
ASME 2"...4", |
JIS 10K |
Material process membrane | 316L, AlloyC, |
Gold-Rhodium |
PE, FEP |
Material gasket | Viton, EPDM, Kalrez, none |
Fill fluid | Inert oil, |
Synthetic oil |
Material housing | 316L, Die-cast aluminum |
Communication | 4...20 mA HART |
PROFIBUS PA |
FOUNDATION Fieldbus |
Certificates / Approvals | ATEX, FM, CSA, CSA C/US, IEC Ex, TIIS, INMETRO, NEPSI |
|
Safety approvals | Overfill prevention WHG |
SIL |
|
Design approvals | EN10204-3.1 |
NACE MR0175 |
|
Marine approvals | GL/ ABS/ LR/ BV/ DNV |
|
Drinking water approvals | KTW/ NSF/ ACS |
|
Specialities | Modularity to differential pressure and process pressure devices (replacable display, universal electronics) |
|
Diagnostic functions |
|
Hermetically sealed Contite measuring cell (condensate-resistant and climatic-proofed) |